EZPAP
Indications
- Used for the treatment or prevention of atelectasis by increasing FRC (Functional Residual Capacity)
Usage
- Augments lung expansion via expiratory flow resistance
- The goal range for the pressure 10-15 cmH20
- The max pressure should not exceed 20 cmH20. If the pressure is reaching higher decrease the flow
- Give Three cycles of 10 breaths, but can be split up based on patient tolerance. Encourage a cough in between sets
- DO NOT deliver nebulizer treatments in-line with EzPAP